Glassware washing and autoclaving
Autoclave and Glassware clean up protocol
Autoclave code: ask the lab manager. Glassware cycle: GGLab1
Contents
Section titled “Contents”1. Washing glassware
Section titled “1. Washing glassware”- Rinse all glassware at least once before taking it to the glassware washer.
- Make sure to remove any old autoclave tape that might still be on the glassware.
- Put all caps, metal spatulas, and stir bars into the metal cage container.
- Place the glassware in the washer making sure to use the correct adapters based on size.
- If the glassware is too much for a single run you may use the second glassware washer located in the other autoclave room.
- For large mouth glassware and bottle caps use the large flat adapter.
- For all the other glassware you can use which ever adapter works best.
- Run glassware cycle GGLab1; this cycle should run for about 50min.
- At times the glassware washer its tank full of water and depending when the error occurred it might need to be rerun.
- If the glassware washer gives this tank full error you may simply run the standby cycle by switching the washer to standby mode it’s a switch behind the black cover. Make sure to close the cover so it can do the standby cycle should take like 3 minutes.
2. Autoclaving
Section titled “2. Autoclaving”- After the cycle is over cover the glassware with its respective cap if it has one, make sure to cap it loosely. If it does not have a cap use aluminum foil to cover it. Place autoclave tape on each individual item to be autoclaved.
- Place the stir bars and metal spatulas into a plastic container and cover with foil.
- Plastic ware such as plastic containers and graduated cylinders does not need to be sterilized unless asked.
- 4L flasks will be hand washed with soap by the person who uses it before being placed into the shelf.
- The autoclave cycle is the gravity with a 45 minute sterilization time and 30 min dry time when using the large and small bin.
- When autoclaving the large bin only, the gravity cycle can be 30 minute sterilization and 20 min dry time.
- After is done place the glassware back into the shelves.
3. Things to check
Section titled “3. Things to check”Something to make sure to check:
- Make sure every water bath has adequate water they tend to need water added to them every 2 weeks or so.
- For the pipet tips you can refill and autoclave them at the same time as the glassware.