Microscope Checklist
A brief checklist to help you get going on the microscope.
Contents
Section titled “Contents”1. Before Your First Flowcell
Section titled “1. Before Your First Flowcell”1.1 Setup
Section titled “1.1 Setup”- Make sure the hard drive has sufficient space to store the awesome data you’re going to produce today.
- Turn on:
- laser
- microscope
- microfluidics
- Wash microfluidics and injection valves with:
- 0.1M NaOH
- 0.1M HCl
- 5M NaCl (optional; removes residual YOYO-1, if needed)
- ddH2O
- **Carefully** screw the 60x objective into the turret.
- **Carefully** remove prism from MeOH, rinse with MeOH and dry with compressed nitrogen.
- Place a small drop of water on the objective and use an old, scratched slide to find proper z-location.
- Check to make sure the microscope is set to R100.
1.2 After Each Flowcell
Section titled “1.2 After Each Flowcell”- Don’t forget to add a bit more water to 60x WI objective.
- Add a tiny drop of additional oil between the prism and flowcell.
2. After Your Last Flowcell
Section titled “2. After Your Last Flowcell”- Remove the prism, rinse with methanol and store in falcon tube with methanol.
- **Carefully** remove the objective, drag-n-drop clean with optical-grade methanol and lens tissue, and put in objective case.
- Replace the objective cap on the turret.
- Cover microscope with dust cover.
- Wash all tubing and loops with 0.1M NaOH, 0.1M HCl, H2O.
- Turn off microscope, syringe pump and laser.
- Make sure you’ve removed and disposed all syringes and wrappers.
- Empty liquid waste container, if needed.