Octet BLI: Setup and Data Analysis

How to set up, run and analyse a kinetics experiment on the lab’s Octet biolayer interferometer. Written as training material, using antibody (IgG) binding to a protein as the worked example; swap in your own ligand, analyte and step times. One run takes about 40 min.

  1. Materials
  2. Build the method
  3. Load sensors and plate
  4. Run the experiment
  5. Data analysis
  6. References
  • HBS-EP+ pH 7.5, from the 10x stock: 10 mM HEPES, 150 mM NaCl, 3 mM EDTA, 0.05 % Tween 20
  • Add 1 mg/mL BSA. The BSA is in the -20 °C freezer, second shelf, solo box, as 500 µL aliquots of 100 mg/mL.
  • Biosensors appropriate to your capture chemistry. New tip trays are in the drawers underneath the table.
  • Black buffer plate and the green sensor tray, inside the instrument.
  • Sample plate.
  • The Octet-specific multichannel pipette, for picking up sensor tips.
  1. Make the running buffer (section 1.1).
  2. Open the Data Acquisition program (the one with blue dots in its logo).
  3. Open an existing method file, or create a new one.
    1. Most people create a new method every time, so the run’s specifics are recorded without overwriting anyone else’s file.
    2. Methods live under the C: drive, Data, then your own folder.
    3. Choose New Kinetics Experiment, then Basic Kinetics > Blank Experiment.
  1. Set the well conditions for your samples.
  2. Change the buffer icon.
  3. Label each well with its name and molar concentration.
  1. In the top right box, add the steps the assay uses, once each. Edit each well to say what it holds: buffer/baseline, loading, sensor type, and so on.
  2. In the bottom right box, put the steps into the order they run.
Step Time Notes
Baseline 60 s Edit to the type of sensor
IgG capture (“loading”) 100-200 s
Post-capture baseline 60 s
Association 600 s
Dissociation 1200 s

Set the sensor information so the Octet picks up the correct tips. See section 3 for the physical handling.

  1. Use the green arrow at the top right to preview the steps.
  2. Check that every sample and sensor is handled the way you intended.
  1. The tips used by the instrument sit in the green/blue tray inside the Octet. Their positions show which wells of the black buffer plate underneath have already been used.
  2. If there is no space on the green tray or black buffer plate, discard the used tips and the black buffer plate, but keep the green tray.
  3. Load 200 µL of running buffer into each well of the black buffer plate where you will place a sensor tip.
  4. Place the black buffer plate underneath the green tray.
  5. Pick up new sensor chips with the Octet-specific multichannel pipette.
    1. Lift straight up when removing and placing tips.
    2. Do not let the sensors touch the well walls.
    3. Hover closely over the well, then let the tips drop in.
    4. Jostle them gently into place if needed, without disturbing the sensor tip.
  6. Equilibrate the sensor tips in running buffer for 10 min before starting a run.
  1. Check previous runs for the response each loading concentration gave. Response values of 0.5-0.8, and no higher than 1, are ideal (Kamat and Rafique, 2017).
  2. If you do not know the loading concentration, or the concentration of protein to bind, run a concentration series first to find the right response level.
  3. Add 200 µL of sample or buffer to each well.
    1. Remember the buffer-only well, placed after the well holding the first protein or IgG. This well becomes the reference sensor during analysis.
    2. No bubbles.
  4. Load the sample plate into the Octet chamber. A1 is the top right corner.
  5. Make sure the plate is snapped in snugly and cannot move. The Octet shakes the plate during the run, so check this twice.
  1. On the Run Experiment tab, set the file location for the results.
  2. Under Run Settings, check the delay box to hold the start for 10 min and let the plate equilibrate.
    1. You can uncheck shake plate while waiting if the sensors have already soaked for 10 min.
  3. Press the green GO button.
  4. Acknowledge the pop-up asking whether the sensor tips are in buffer.
  1. Open the Data Analysis program from the desktop.
  1. Click your file in the bottom left panel, under OS (C:) > Data > <your name> > <experiment folder>. It then appears in the top left panel.
  2. Open the file from the top left panel.
  3. Ignore the Experiment Summary tab.
  4. In the Sensor-Assay tab, select the tips to analyse.
  1. Step 1: set the buffer-only sensor as the reference. Select the well holding buffer only, right-click, choose Edit Sensor Properties, and change Sensor Type to Reference Sensor.
  2. Step 2: check the Subtraction box and click Average Reference Sensors.
  3. Step 3: click align Y-axis.
  4. Step 4: skip.
  5. Step 5: click Process Data!
  1. Under Curve Fitting, select Association and Dissociation.
  2. Set the model to 1:1.
  3. Choose the fitting mode:
    • Local fits each sensor and its own KD. Use this for cross-reactivity studies.
    • Global fits all sensors to one KD. Use this for a competition assay run at several concentrations of the same sample.
  4. Press Fit Curves.
  5. Screen grab the Align Y-axis graph and copy it into a PowerPoint file.
  1. Click Fit Data to show the fit lines on all the curves.
  2. If the curves and fits are clean, as in a competition assay, take one screen grab showing the fits and save it to the PowerPoint file.
  3. If they are messy, as in a cross-reactivity assay, screen grab each one separately. Toggle between fits by clicking rows in the table.
  4. Read the kinetics values from the large table: KD, kon, and koff, which the software labels Kdiss.
  5. Save the kinetics table as a CSV file.
  • Shah NB, Duncan TM. Bio-layer interferometry for measuring kinetics of protein-protein interactions and allosteric ligand effects. J Vis Exp (2014). doi:10.3791/51383 - setup video and overview of the Octet.
  • Kamat V, Rafique A. Designing binding kinetic assay on the bio-layer interferometry (BLI) biosensor to characterize antibody-antigen interactions. Anal Biochem (2017). doi:10.1016/j.ab.2017.08.002 - ideal response unit range.