FPLC Checklist
Contents
Section titled “Contents”1. Before you begin
Section titled “1. Before you begin”Always filter all FPLC buffers using a 0.45 or smaller µm filter, degas, and chill to 4°C to avoid bubbles and clogging
- Treat your lysate with DNAse and use the Ultracentrifuge to clarify.
- Perform system and pump washes with water to remove ethanol. The salt in purification buffers will crash out when exposed to ethanol and can destroy columns.
- Purge the pumps to remove air from the pump heads.
2. Beginning a purification run
Section titled “2. Beginning a purification run”- Set alarms for pre column and delta column pressures according to the column manual.
- Perform “Pump Wash A & Pump Wash B” with water.
- Wash column with ~10 column volumes (CV) of water. Set pause timer.
- Wash column with ~10 CV of buffer B to flush out residual protein from a previous prep. Set pause timer.
- Wash column with ~10CV of buffer A. Set pause timer.
3. Finishing a purification run
Section titled “3. Finishing a purification run”If running manually, set appropriate time/volume pause timers to avoid running the system dry.
- Wash column with 100% Buffer B to ensure protein is completely eluted.
- Wash column with H2O.
- Store column in water (for up to 48 hours) or 20% EtOH (>48 hours between use) and label the column with the storage solution, the date, and your initials.
- Clean the syringe port with HCl, NaOH, and water.
- Perform a system wash with water and then with 20% EtOH.
- Empty the fraction collector.
- Wipe the dispenser head and sensor diode of the fraction collector with an ethanol-soaked Kimwipe.